An experimental study on the capture effect in 802.11a networks
Proceedings of the second ACM international workshop on Wireless network testbeds, experimental evaluation and characterization
Performance Study of Wireless Body Area Network in Medical Environment
AMS '08 Proceedings of the 2008 Second Asia International Conference on Modelling & Simulation (AMS)
A study of implanted and wearable body sensor networks
KES-AMSTA'08 Proceedings of the 2nd KES International conference on Agent and multi-agent systems: technologies and applications
Hi-index | 0.00 |
In this paper, we implement Implantable Cardioverter Defibrillator (ICD) system. In order to implement the ICD, we propose the baseband according to modulation method, modulation index and bandwidth time product. From simulation results, we decide GFSK modulation with modulation index 0.2 and bandwidth time product 0.5. Also the receiver sensitivity of ICD link budget decides -83.22dBm using receiver antenna gain, receiver power and noise figure. In test and verification of implemented SoC prototype, we obtain the packet error rate 1% at -83dBm.