Verification of Set Up Dual-Chamber Pacemaker Electrical Parameters

  • Authors:
  • Augustynek M.;Labza Z.;Penhaker M.;Korpas D.

  • Affiliations:
  • -;-;-;-

  • Venue:
  • ICCEA '10 Proceedings of the 2010 Second International Conference on Computer Engineering and Applications - Volume 02
  • Year:
  • 2010

Quantified Score

Hi-index 0.00

Visualization

Abstract

The measurement on a dual-chamber cardiostimulator consists of the construction of a unit for the simulation of the cardiac action in a patient’s body. It has an adjustable amplitude (0 - 20 mV), frequency (0 - 100 Hz) and mark-space ratio. The system output impedance should be in the range of 200 - 2000 Ω. It is possible to add a defined noise into the output channel. It is further possible to measure the tolerance of sensing of the atrial and ventricular channel of the dual-chamber cardiostimulator DDDR.