Partial volume correction for positron emission images using the wavelet transform

  • Authors:
  • Antonello E. Spinelli;Carlo Bergamini;Romano Zannoli;Gianfranco Cicoria;Mario Marengo;Luca Guerrieri;Claudio Lamberti

  • Affiliations:
  • Policlinico S. Orsola-Malpighi, Department of Medical Physics, Bologna, Italy;Policlinico S. Orsola-Malpighi, Department of Medical Physics, Bologna, Italy;Policlinico S. Orsola-Malpighi, Department of Medical Physics, Bologna, Italy;Policlinico S. Orsola-Malpighi, Department of Medical Physics, Bologna, Italy;Policlinico S. Orsola-Malpighi, Department of Medical Physics, Bologna, Italy;University of Bologna, DEIS, Bologna, Italy;University of Bologna, DEIS, Bologna, Italy

  • Venue:
  • WAMUS'06 Proceedings of the 6th WSEAS international conference on Wavelet analysis & multirate systems
  • Year:
  • 2006

Quantified Score

Hi-index 0.00

Visualization

Abstract

A method for partial volume correction of positron emission tomography (PET) images is proposed. The method is based on combining information from PET and high resolution anatomical images. Two different algorithms fast wavelet transform and à trous were implemented in order to decompose the PET and high resolution images. Results obtained using several simulated images show a significant partial volume reduction and image enhancement. The relative errors in the partial volume corrected image are always less than 3, 6% with respect to 16% of the original image.