Establishing calibration models and error envelopes of a 3D whole body scanning system

  • Authors:
  • Yu-Cheng Lin;Mao-Jiun J. Wang;Toly Chen

  • Affiliations:
  • Department of Industrial Engineering and Management, The Overseas Chinese Institute of Technology, Taichung City, Taiwan, R.O.C.;Department of Industrial Engineering and Engineering Management, National Tsing Hua University;Department of Industrial Engineering and Systems Management, Feng Chia University

  • Venue:
  • IMCAS'06 Proceedings of the 5th WSEAS international conference on Instrumentation, measurement, circuits and systems
  • Year:
  • 2006

Quantified Score

Hi-index 0.00

Visualization

Abstract

The accuracy is important for a 3D whole body scanning system. In order to analyze the measurement error, a set of dimension-certified measurement equipment whose true dimensions were certified was introduced. The calibration models of height, breadth and depth measurement between measured values and true dimensions were established respectively and relevant error envelopes were derived. A verification experiment was conducted to determine the validity of these models. The accuracy of this scanning system can be improved by using the calibration models to obtain more accurate 3D human models.