An asymmetric dialysis tube as an integrated part of a pressure-monitoring sensor

  • Authors:
  • Roger Andersson;Per Ask

  • Affiliations:
  • -;-

  • Venue:
  • Technology and Health Care
  • Year:
  • 2001

Quantified Score

Hi-index 0.00

Visualization

Abstract

A sensor has been designed consisting of a tube holder with a forcetransducer and a tube with a modified cross-section. The holder hasa lid that encloses the tube. By having a stiff holder and acompliant tube, the idea is that the intraluminal pressure in thetube can be obtained from the measured force. The method isintended for non-invasive pressure measurements in blood ordialysate tubes. We have used a tube cross-sectional geometry where the outersurface is elliptic and the inner surface is circular with arelation of 2:1 between the thinnest and thickest tube sides. Thepressure transducer system shows a linear relationship between theapplied pressure and the sensor output (r=0.999). Within thetemperature range, 32°--36°C, which corresponds to theblood and dialysate temperatures, the sensor accuracy is within± 0.8 kPa (± 6 mm Hg). This indicates that the sensorshould be clinically useful during dialysis and similarapplications.