2D-3D registration of CT vertebra volume to fluoroscopy projection: a calibrationmodel assessment

  • Authors:
  • P. Bifulco;M. Cesarelli;R. Allen;M. Romano;A. Fratini;G. Pasquariello

  • Affiliations:
  • Department of Biomedical, Electronic and Telecommunication Engineering, University of Naples Federico II, Napoli, Italy;Department of Biomedical, Electronic and Telecommunication Engineering, University of Naples Federico II, Napoli, Italy;Institute of Sound and Vibration Research, University of Southampton, Southampton, UK;Department of Biomedical, Electronic and Telecommunication Engineering, University of Naples Federico II, Napoli, Italy;Department of Biomedical, Electronic and Telecommunication Engineering, University of Naples Federico II, Napoli, Italy;Department of Biomedical, Electronic and Telecommunication Engineering, University of Naples Federico II, Napoli, Italy

  • Venue:
  • EURASIP Journal on Advances in Signal Processing - Image processing and analysis in biomechanics
  • Year:
  • 2010

Quantified Score

Hi-index 0.00

Visualization

Abstract

This study extends a previous research concerning intervertebral motion registration by means of 2D dynamic fluoroscopy to obtain a more comprehensive 3D description of vertebral kinematics. The problem of estimating the 3D rigid pose of a CT volume of a vertebra from its 2D X-ray fluoroscopy projection is addressed. 2D-3D registration is obtained maximising a measure of similarity between Digitally Reconstructed Radiographs (obtained from the CT volume) and real fluoroscopic projection. X-ray energy correction was performed. To assess the method a calibration model was realised a sheep dry vertebra was rigidly fixed to a frame of reference including metallic markers. Accurate measurement of 3D orientation was obtained via single-camera calibration of the markers and held as true 3D vertebra position; then, vertebra 3D pose was estimated and results compared. Error analysis revealed accuracy of the order of 0.1 degree for the rotation angles of about 1mm for displacements parallel to the fluoroscopic plane, and of order of 10mm for the orthogonal displacement.